Error Of The Channel - ProMinent PROFIBUS DP DULCOMETER diaLog DACb Supplementary Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

Supplementary instructions PROFIBUS

1.6.2 Error of the channel

Bit
Description
31
Error 99: There is a system error;
30
29
28
27
26
25
24
23
22
21
20
Error 88: The connection to the extension module is faulty;
[The connection to the expansion module is faulty ]
19
Error 34: Incorrect correction variable;
18
Error 19: The liquid level in storage tank 3 is too low;
17
Error 18: The liquid level in storage tank 2 is too low;
16
Error 17: The liquid level in storage tank 1 is too low;
15
Error 16: The mA input is overloaded;
14
Error 15: The mA input supply is overloaded;
13
Error 14: The status of the controller is pause / hold
[The controller is in the state PAUSE / HOLD]
12
Error 13: The status of the controller is pause
11
Error 12: There is a sample water fault e.g. no flow;
10
Error 11: After elapse of the delay period, a limit value error still exists;
[After elapsing of the delay time a limit error still exists]
9
Error 10: The mA input current is less than 4 mA;
8
Error 9: The mA input current is greater than 20 mA;
[The mA input current is greater than 20 mA ]
7
Error 8: The check time was infringed;
6
Error 7: Check the mechanical condition (glass breakage) of the sensor;
[Check the mechanical status of the sensor Glass break is possible]
5
Error 6: No sensor available;
4
Error 5: Calibration error exists;
3
Error 4: The temperature is too high;
2
Error 3: The temperature is too low;
1
Error 2: The mV input voltage is too high;
0
Error 1: The mV input voltage is too low;
18
®
DP
[A system error exists]
[Incorrect correction variable ]
[The level in tank 3 is too low ]
[The level in tank 2 is too low ]
[The level in tank 1 is too low ]
[The mA input is overloaded]
[The mA input supply is overloaded]
[PAUSE / HOLD] ;
[PAUSE] ; [The controller is in the state PAUSE]
[Error sample water exists, e. g. no flow]
[The mA input current is less than 4 mA ]
[The checkout time was infringed]
[No sensor is available ]
[A calibration error exists]
[The temperature is too high]
[The temperature is too low ]
[The mV input voltage is too high ]
[The mV input voltage is too low ]