15.1.7. Install the plastic calibration cups into the digestion vessel guide, ensuring that the lip of the cup is flush with
the guide and that the pour spout is facing towards the user.
15.1.8.
SAFETY ALERT:
press "Start Test" until fully
completing the instructions below.
Close the digestion oven door before
starting the test. Press "Start Test." The
lines will be primed, and solutions
delivered. The test will pump 10ml volume
to each cup sequentially.
There are two faults that may occur during the solution delivery calibration:
Empty Supply Line
This fault is most often caused by an empty reservoir. The user
must refill the reservoir and press continue.
If the message continues to be displayed, it may be caused by a
restriction in the line or a leak in the system. Contact ANKOM
Technology for assistance.
Full Supply Line
This message can occur if the lines are not cleared properly at the end of the solution delivery calibration.
Possible causes include left over liquid in the line, full waste container, or mechanical failure. Contact
ANKOM Technology.
pg. 46
Do not
Digestion Vessel Guide
Rev 1/22/2021