Preventative Maintenance Process Guide
Date: _________________________
Description:
This Process Guide provides instructions to complete the Preventative Maintenance
(PM) process for the ANKOM DELTA or the A2000 Automated Fiber Analyzer.
This process guide assumes there are no current faults or leaks within the instrument.
Frequency:
The Preventative Maintenance process is recommended to be completed annually for
instruments used daily or every 2—3 years for lesser used instruments.
Supplies:
DELTA/A2000 Preventative Maintenance Kit (Part #: DELTA20)
Preparation
This phase includes gathering the needed parts and tools.
1. Kit Contents:
DELTA/ A2000 Preventative Maintenance Kit (DELTA20): Includes parts listed below appropriate
for either the ANKOM DELTA or A2000 Automated Fiber Analyzer:
•
F26—Packing Kit
•
640—Brass Packing Nut
•
632—Helical Coupler
•
F11.5—Bag Suspender Kynar Tip
•
Z541—Packing Installation Tool
2. Gather the following tools:
•
Hex (Allen) driver 1/8" (Included with instrument)
•
1
" Open-end wrench (Included with Instrument)
1/8
•
Small flat blade (standard) screwdriver
•
1/2" or 13mm Open-end wrench
•
7/64" Allen Wrench
Prechecks Before You Start:
signs of leakage from the vessel or the motor assembly. If signs of leaking are observed from the mo-
tor, it is recommended that the motor assembly be replaced (A08 for 110v instruments or A10 for
220v instruments) as part of the Preventative Maintenance. Pause the Preventative Maintenance and
obtain the appropriate motor for the instrument configuration if leaking has occurred.
Also verify the efficiency of the level sensor by entering diagnostics and inspecting the level sensor
reading (refer to Level Sensor Check procedure on pages 6 or 7 in this process guide). If the value is 5
or less, clean the glass tip on the level sensor probe in the vessel and check the reading again. Use a
cotton swab with alcohol to wipe the tip of the level sensor. Cleaning Procedure can be found at
https://www.ankom.com/sites/default/files/document-files/
DES007_Level_Sensor_Analog_Cleaning_and_Testing.pdf
If the value is still 5 or less, replace the level sensor assembly (A32) before starting the Preventative
Maintenance process.
ANKOM Technology • 2052 O'Neil Road • www.ankom.com/contact/technical-services • Phone: 315-986-8090 • Fax: 315-986-8091
Instrument Serial # __________________________________
It is advisable to inspect the base of the instrument looking for
page 1 of 7
-
DELTA/A2000
DES042
Service Procedure