Operator's Manual
6.9.2.2. Wet the bottom outside edge of the
digestion vessel with water. Do not use any
other solution, as this will negatively impact
the seal.
6.9.2.3. Insert the digestion vessel into the vessel
bottom assembly by a slight twist motion.
IMPORTANT:
the digestion vessel seats evenly
and tightly, with the glass all the
way down in the vessel bottom
assembly, otherwise the digestion
solution will leak.
6.9.3. Load Method: Line Prime
From the Home Screen:
6.9.3.1. Press "METHODS."
6.9.3.2. Select the "MISCELLANEOUS" category.
6.9.3.3. Select the "Line Prime" method.
6.9.3.4. The screen will display the steps required to start
a run. Press the first step, "1. Enter sample
name."
pg. 20
Make sure
Rev 1/22/2021