13. Periodic Maintenance

Inspect and clean solution reservoir bottles and filters

On a monthly basis (or more frequently depending on usage), follow the steps below to inspect and clean the solution reservoir
bottles and filters.
13.1.1. Inspect the bottles, filters, and internal tubing, for any precipitate, particles, or foreign matter.
13.1.2. Clean filters by running them under fast running water. Replace as needed.
13.1.3. Wash and rinse the bottles out with water and let air dry as needed.

Inspect and clean the spray nozzles in the digestion oven

On a monthly basis (or more frequently depending on usage), follow the steps below to inspect and clean the spray nozzles.
13.2.1. Periodically observe the spray capability during the addition of solution in the digestion vessels.
13.2.2. Clean spray nozzles by wiping each nozzle with a wetted cloth or paper towel.

Regularly clean the instrument surfaces

13.3.1. Wipe down the surface of the instrument whenever spills have occurred.