Nhs-35 Jet Assembly - Agilent Technologies HS-16 User Manual

High throughput diffusion pump
Hide thumbs Also See for HS-16:
Table of Contents

Advertisement

Technical Information

NHS-35 Jet Assembly

Figure 29 NHS-35 Jet Assembly
To disassemble the jet:
1
Unscrew and remove the hex nut holding the cold cap in place. Note its
orientation prior to removal
2
Remove the cold cap (or halo baffle) as described in "Cold Cap".
3
Unscrew the top jet cap and remove.
4
Remove the spacer and top jet orifice plug.
5
Lift the second stage and jet shield over the center threaded jet rod.
6
Using the two eyebolts now accessible and appropriate lifting equipment, lift
the entire lower jet assembly out of the pump body, leaving the jet rod in place.
7
Remove the third stage jet shield.
8
Remove the securing nuts and lift the third stage and fractionating tube off the
fourth stage.
9
Remove the fourth state jet shield.
10 Remove the securing nuts and separate the fourth stage from the jet base.
11 Clean and service as needed.