Complete cleaning of the pump may be required due to gradual deterioration of
pump fluids. Removal of the pump from the system is then necessary.
To clean an installed pump:
1
Disconnect all water cooling lines and break the primary circuit supplying power
to the pump heaters.
2
Unbolt the inlet and foreline connections and remove the pump from the
system.
3
Drain the pump of all fluid.
4
Remove the cold cap as described in "Cold Cap".
5
Remove the jet assembly as described for the specific pump model in "Jet
Assemblies".
6
Thoroughly clean the pump body interior using acetone followed by an
isopropyl alcohol rinse and then dry the pump with clean, dry, oil-free
compressed air.
7
Thoroughly clean the jet assembly using acetone. Wipe all surfaces with
isopropyl alcohol, and dry with clean, dry, oil-free compressed air.
8
Reinstall the jet and cold cap in the pump body. Be sure that the ejector is
properly aligned with the foreline.
9
Check the condition of the O-rings. Replace O-rings that show any sign of wear
or compression.
10 Attach the pump in the system.