Technical Information
Disassembly and Reassembly Procedures
Cold Cap
To disassemble the cold cap, refer to the following figure and take the following
steps.
The halo baffle is disassembled in the same manner.
NOTE
1
Remove the female coupling, nut, follower, and gasket located at the end of the
cold cap water line on the outside of the pump.
2
Remove the cold cap securing screw which secures the cold cap to the jet
assembly.
3
Lift out the cold cap.
To reassemble the cold cap:
1
Insert the end of the cold cap water line into the cold cap port opening at the
side of the pump before setting the cold cap on top of the jet assembly. Be
careful not to damage the sealing surfaces.
2
Replace the cold cap securing screw which secures the cold cap to the jet
assembly. Do not overtighten the screw.
3
Replace the gasket, the follower, the nut, and the female coupling to the side of
the pump.
62
High Throughput Diffusion Pumps User Manual