Dräger Fabius 4116029 Technical & Service Manual page 88

Anesthesia system
Table of Contents

Advertisement

Repair Instructions
Ventilation frequency or inspiratory
51
time out of tolerance.
RAM (Power PCB)
81
EPROM (Power PCB)
82
Watchdog reset (Power PCB)
83
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Motor does not meet
specifications.
Ventilator needs replaced.
Error may also occur if Fabius
is operated while ventilator is
not connected.
RAM needs replaced.
EPROM needs replaced.
Software error. Duration
exceeded.
3-20
®