10.7
Front Panel
The front panel consists of the front frame and the membrane keypad.
10.7.1
Membrane Keypad
The membrane keypad has 15 keys (man/spont, IPPV, Pmax, VT, f
PEEP, upper and lower limit value setting, flow calibration, silence, standby, and 3
softkeys). The Control PCB scans the key operations.
10.7.2
Man/Spont Key
Fabius can be set to manual ventilation or spontaneous breathing by pressing the
Man/Spont key.
10.7.3
IPPV Key
Fabius can be set to intermittent positive pressure ventilation by pressing the IPPV
key.
10.7.4
Pmax Key
The inspiratory pressure limit can be set in a range of 10 hPa (mbar) to 70 hPa (mbar)
by pressing the Pmax key. A pressure higher than 40 hPa (mbar) must be confirmed
by pressing the Reset/Check key (a message is shown on the display).
10.7.5
V
Key
T
After pressing the V
10.7.6
T
:T
Key
I
E
After pressing the T
of 1:3 to 2:1. The set value becomes effective at the end of the current respiratory
cycle.
10.7.7
T
:T
Key
IP
I
After pressing the T
in a range of 5% to 50%. The set value becomes effective at the end of the current
respiratory cycle.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
key, the tidal volume can be set in a range of 50 mL to 1400 mL.
T
:T
key, the inspiratory/expiratory time ratio can be set in a range
I
E
:T
key, the inspiratory pause time to inspiratory time can be set
ip
i
2-55
RETURN TO CD-ROM TABLE OF CONTENTS
Function Description
IPPV
, T
:T
T
:T
I
E,
IP
I,