Repair Instructions
23.1
Dismounting/Replacing the Patient System
• Unlock patient system 1 by turning screws 2 one quarter of a full rotation (90°)
counterclockwise using an Allen key.
• Lift and remove patient system 1.
2
Fig.159:Removing the patient system
• Place new patient system 1 onto ventilator.
• Secure patient system 1 by turning screws 2 one quarter of a full rotation (90°)
clockwise using an Allen key.
23.2
Safety Valve
In case of repair, replace the complete safety valve.
23.2.1
Testing the Safety Valve
• Check function of safety valve according to Test Certificate.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
2
1
3-162
®
Fabius