21.6
Pressure-limiting valve test
On control box IPPV, set
p
= 40mbar
max
f
= 6/min
IPPV
TI:TE
= 1 : 1
TIP:TI
= 10 %
PEEP
=0mbar
V
= 380 mL
t
Connecting sleeve M 13506
Pressure gauge 7901515
Connect -30/+120 mbar pressure gauge
to ventilator outlet.
Relief valve opens at 75 ± 5 mbar
(maximum pressure)
21.7
Testing of ancillary air valve:
Flow control valves on 0. Vacuum valve
opens at -2.5 to -6 mbar (minimum
pressure)
21.8
Freedom from leaks in piston chamber
Select Man/Spont mode.
Connect flowmeter and pressure gauge to
ventilator connection. Piston in upper end
position.