Fabius
6.3
Debug Mode Table (software 1.00)
Table 3: Debug Mode (software 1.00)
Pressure sensor value (Paw) in mbar
(default)
P
XXX (0 to 738)
Flow sensor A/D converter value (default)
FL
XXX (0 L/min 512)
Oxygen sensor channel 1 in mV
O2_1
XXX (9 – 20)
Oxygen sensor channel 2 in mV
O2_2
XXX (9 – 20)
Expiratory Vt during inspiration (mL)
Vtix
XXX (varies with system settings,
ideally 0)
Pressure sensor calibration value
CalP
XXX (704 – 768)
Flow sensor calibration value
CalF
XXX (482 – 542)
Oxygen sensor calibration value (21% O2)
CalO
XX (39 – 113)
Monitoring error code
MONI
XXXX
Calculation from controller data (set values):
Tidal volume (mL)
V:
XXXX
Error log
(most recent error entry on extreme left)
LOG
XX XX XX XX XX XX XX XX XX
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Plateau pressure in mbar
Plat
settings)
Supply voltage (approx. 27 V) in volts
U_24
Supply voltage (+12 V) in volts
U+12
Supply voltage (–12 V) in volts
U–12
EEPROM error code
EEPR
Battery charging current in amperes
LADI
Motor current in amperes
MOTI
Pressure between rolling seals in mbar
PMEM
Ventilation error
VENT
System temperature in °C
frequency (1/min)
F:
XX
T
3-9
Repair Instructions
XXX (varies with system
XX.X (22.5 – 35.0)
XX.X (12.0 ±2.0)
XX.X (12.0 ± 2.0)
XXXX
X.X (approx. 0.0 – 2.0)
X.X (0.0 – 2.5)
XXX (200 ±50)
XXXX
XX (max. 60°C)