Function Description
12
Flow Measurement
A moving screen (moving vane) measures the expiratory flow of the patient. A
flow-dependent pressure builds up at the moving vane. This pressure is picked up at
the differential pressure sensor ( ∆ P). The differential pressure sensor converts the
pressure into an electronic signal and transmits this signal to the microcontroller.
.
V
measuring line
P
measuring line
Fig. 41: Flow measurement principle
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
P
flow sensor
moving vane
P results from the differential pressure
(back-pressure from measuring lines 1 and 2)
2-58
®
Fabius