• Hold electronic PEEP valve 1 in place and remove fixing screws and washers 2.
Fig.156:Top view of the pneumatics, replacing the electronic PEEP valve
• Apply Loctite #222 on threads of fixing screws 2 and mount new electronic PEEP
valve to pneumatics.
• Connect hoses to connection ports 3 of electronic PEEP valve (according to
previously applied markings).
• Connect connector 4 of electronic PEEP valve to connection 5 on
Pneumatics Control PCB.
• Fasten all electrical cables together using a cable tie.
• Cut off any protruding length of cable tie.
• Mount pneumatics to Fabius using the reverse method of that used for removal.
• Connect gas supplies and electrical connections to Fabius.
• Connect complete breathing system (breathing hoses, Y-piece, breathing bag
etc.) to circle absorption system.