Repair Instructions
• Disconnect connector 1.
• Mark hoses 2 and 3 for later reassembly and disconnect them from piston pump
connection.
Fig.146:Top view of the pneumatics, removing the piston pump
• Carefully cut cable tie which fastens electrical cables.
• Push rubber mounts 4 upwards and out through pneumatics plate 5.
• Mount rubber mounts 4 to new piston pump.
• Mount new piston pump to pneumatics plate 5 by carefully pushing rubber
mounts 4 through pneumatics plate 5 using a screw driver.
4
Fig.147:Bottom view of the pneumatics, removing the piston pump