Dräger Fabius 4116029 Technical & Service Manual page 145

Anesthesia system
Table of Contents

Advertisement

®
Fabius
14.19.3 Dismounting/Replacing the Backlighting Lamp
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Remove control box from Gas Box (see "Removing the Control Box from the Gas
Box").
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
• Remove lateral screws 1 from control box 2 and lift top cover 3 in direction of A
and away from control box 2.
• Place top cover 3 of control box 2 aside.
1
Fig. 60: Front view of control box, removing top cover
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
A
IPPV
Paw [mbar]
24
80
FiO2
60
21
7.9
40
12.0
MV
man./
3.0
20
spont.
28
40
Paw
0
8
IPPV
Pmax
V
f
T
I
:T
E
T
IP
:T
I
PEEP
T
IPPV
3-77
RETURN TO CD-ROM TABLE OF CONTENTS
Repair Instructions
Menu
1..2..
0
3
1
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents