Dräger Fabius 4116029 Technical & Service Manual page 249

Anesthesia system
Table of Contents

Advertisement

11.1.1.1
Sealing ring
11.1.2
Renew bacteria filter CH 102
*
(set of 5: 6723976)
11.2
Vacuum-type secretion aspirator
assembly
11.2.1
Vacuum connecting hose
11.2.1.1
Sealing ring
11.3
Frame
11.3.1
Cylinder holder
11.3.2
Connection hose
11.3.3
Secretion aspiration hose M 25780
11.3.4
Secretion sight glass/fingertip
11.4
Jar M 20091, 2x
11.5
Jar cap
11.5.1
Check tightness of collar M 26008 (renew
*
if applicable).
11.6
Float M 26007
*
(renew if applicable - set of 5: M 24733)
11.6.1
Freedom of movement
To check:
Connect up aspirator assembly
Build up vacuum
Turn jar round
Float is sucked in
Turn jar round
Shut off vacuum
Float drops back
11.7
Relief valve
(if fitted)
11.7.1
Mica insulator R 17329
(renew if applicable - set of 10: R 23192)
11.8
Ejector
Connect up test vacuum gauge via
secretion jar.
Capacity: p = > -0.8 bar
Of which: p = -0.7 bar in 10 s