Dräger Fabius 4116029 Technical & Service Manual page 184

Anesthesia system
Table of Contents

Advertisement

Repair Instructions
19.2.7
Fine Adjusting O
i
Make sure O
calibrated. Check O
1 vol.% O
• Fully open N
O flow control valve on flowmeter block.
2
• Adjust O
flow control valve to 1 L/min.
2
N
O flow should be 3 L/min (±0.3 L/min). O
2
be 25 vol.% O
(±3 vol.% O
2
• If N
O flow is not 3 L/min (±0.3 L/min) or if O
2
not 25 vol.% O
(±3 vol.% O
2
• Use adjusting screw 1 on N
A
0
2
Fig.104:S-ORC, adjusting the N
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Concentration (O
2
2
sensor is functioning properly and has been
2
concentration using an O
2
).
2
concentration at fresh-gas outlet should
2
).
2
), proceed as follows:
2
O side to adjust above mentioned values.
2
E
1
N
0
2
A
E
O flow
2
3-116
RETURN TO CD-ROM TABLE OF CONTENTS
/N
O ratio)
2
analyzer (accuracy:
2
concentration at fresh-gas outlet is
2
®
Fabius