Repair Instructions
• Remove lateral screws 1 from control box 2 and lift top cover 3 in direction of A
and away from control box 2.
• Place top cover 3 of control box 2 aside.
1
man./
spont.
IPPV
Fig. 22: Front view of control box, removing top cover
• Check that fuses on Power Supply PCB are OK.
If fuses on Power Supply PCB are not OK, proceed as follows:
• Plug control box power cord into AC outlet.
• Switch control box on.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
A
IPPV
Paw [mbar]
24
FiO2
80
60
21
40
7.9
12.0
MV
3.0
20
28
40
1..2..
Paw
0
8
Pmax
V
f
T
:T
T
:T
PEEP
T
IPPV
I
E
IP
I
0
3-38