14.14.3 Dismounting/Replacing the Pressure Sensor (Flow)
i
Two different types of differential pressure sensors for flow
measurement may exist in the field. Fabius CE printed circuit boards
are designed for either type. However, we recommend using the type
manufactured by Data Instruments.
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Remove control box from Gas Box (see "Removing the Control Box from the Gas
Box").
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
• Remove lateral screws 1 from control box 2 and lift top cover 3 in direction of A
and away from control box 2.
• Place top cover 3 of control box 2 aside.
1
man./
spont.
IPPV
Fig. 37: Front view of control box, removing top cover
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
A
IPPV
Paw [mbar]
24
FiO2
80
60
21
7.9
40
12.0
MV
3.0
20
28
40
1..2..
Paw
0
8
Pmax
V
f
T
:T
T
:T
PEEP
0
T
IPPV
I
E
IP
I
3-56