Connecting the sensor cables

When connecting the sensor to the DSL make sure that you connect the sensor cables for each
path to the corresponding numbers on the DSL (A and B).
External DSL side A
The sensor cables are delivered with a pre-mounted F-connector at transmitter end.
1. Remove blind plug from external DSL enclosure.
2. Slide the cable gland back onto the cable and allow access for the F-connector tool.
3. Place F-connector tool around cable and slide up to engage F-connector nut.
4. Push cable through gland opening. Ensure that center lead is aligned with connector port in
external DSL.
5. Tighten F-connector using tool until you meet mechanical stop.
6. Remove F-connector tool.
7. Mount and tighten cable gland.
Repeat these steps for each sensor cable.
SITRANS FS230
Compact Operating Instructions, 05/2021, A5E38755607-AC