Cables And Cable Entries - Siemens SITRANS FS230 Compact Operating Instructions

Ultrasonic flowmeters
Table of Contents

Advertisement

Technical specifications

7.10 Cables and cable entries

7.10
Cables and cable entries
Table 7-30
Description
Number of conductors
Square area [mm
Screen
Outside color
External diameter [mm]
Maximum length [m (ft.)]
Installation environment
Insulation material
Halogen-free
RoHS compliant
Torsional strength
Permissible temperature range [°C (°F)]
Min. bending radius allowed
Table 7-31
Description
Square area [mm
Linear resistance [Ω/km]
Max. length [Ω] (depends on total linear resistance) < 500
Signal run time [ns/m]
Insulation resistance [MΩ*km]
Characteristic impedance 1 – 100 MHz [Ω]
Attenuation @ 1 MHz
Operating voltage (peak) [V]
Test voltage (wire/wire/screen rms 50 Hz 1 min) [V] = 700
Electrical data at reference temperature (20 °C)
Table 7-32
Description
Square area [mm
Max. length [m]
104
SSL cable, basic data
]
2
Signal cable recommendations for channels 1 to 4
]
2
Power supply cable recommendations
2
]
Specification
4
0.326 (AWG 22/7)
Yes
Standard version: gray (RAL 7001)
Ex version: light-blue (RAL 5015)
6.5 (standard); 12 (armored)
75 (246)
Industrial including chemical processing plants
Special polyolefin
Yes
Yes
>3 million cycles at ± 180° on 200 mm
Not adapted for garland mounting (festoon)
-40 to +80 (-40 to +176)
Single 5 X ø
Specification
0.5 (AWG 20)
≤ 120
≤ 5.3
≥ 200
100 (±5)
< 2.9 dB/100 m
≤ 300
Specification
1.3 (AWG 16)
300 (AWG 16)
Compact Operating Instructions, 05/2021, A5E38755607-AC
SITRANS FS230

Advertisement

Table of Contents
loading

Table of Contents