Connecting The Sensor Cables (Internal Dsl Only) - Siemens SITRANS FS230 Compact Operating Instructions

Ultrasonic flowmeters
Table of Contents

Advertisement

Connecting
4.7 Connecting wall mount enclosure
For configuration of the software parameters, see Input/output configuration (Page 73). More
information can be found in the Function Manual.
The following table shows:
• Which cable with which terminal
• Hardware and software configuration of the channels
HW configuration
SW configuration
Power supply
Channel 1 HART
Current output HART
Channel 1 MODBUS
Channel 1 PROFIBUS
Channel 2 output
Current, Frequency, Pulse and Status
Outputs: Current, Frequency, Pulse and Status
Channel 3 & 4 Input and output
Inputs: Digital
Channel 3 & 4 Relay
Status output
Figure 4-11
Terminals/configuration overview
Ex and non-Ex versions
• For Ex versions active or passive current output is preselected at the time of ordering and
cannot be changed.
• Non-Ex versions can be connected as either active or passive.
4.7.2

Connecting the sensor cables (internal DSL only)

Preparing for the connection
1. Loosen the four lid screws.
2. Open the lid.
3. Remove F-connector tool from wall mount enclosure.
Figure 4-12
64
Power supply
Ch1
1
2
3
4
5
L/+
N/-
Active/+
Active/ -
Passive/+
In
+ (B)
- (A)
In
+ (B)
- (A)
F-connector tool location
Terminals
Ch2
6
7
8
9
10
Passive/-
Out
+ (B)
- (A)
Out
+ (B)
- (A)
Active/+
Active/ -
Passive/+ Passive/-
Compact Operating Instructions, 05/2021, A5E38755607-AC
Ch3
Ch4
11
12
13
14
15
Active/+
Active/ -
Active/+
Active/ -
Passive/+
Passive/-
Passive/+ Passive/-
NO
NO
NO
NO
NC
NC
NC
SITRANS FS230
16
NC