The screen displays the first contaminant type and size to be used.
2. Select the required sample test pack and pass it through the system the required
number of times until the
Each test pack should display as being rejected by the system which is correct.
If a pack passes inspection then effectively the test has failed for that test
sample and the test should be cancelled. The product inspection settings
should be reviewed and possibly another product learn process carried
out.
If any test fails
Any product packs that have passed inspection since the last successful PVS Test was
carried out should then be re-inspected.