10. Calibration is complete. The ANSI values of the barcodes now match the recorded values on the card within
a few points (+/- 8%). If calibration fails, repeat the calibration process. Be sure to inspect the calibration card
carefully for damages and impurities. If the problem still persists, contact Omron Microscan for technical
assistance.
11. Click the "Save calibration and exit" button to save the calibration results and return to the Welcome
screen. After saving the calibration results, the system automatically saves the Operator ID in the calibration
history log (see the section entitled "Review Calibration History" for more information). Click "Discard
changes (if any) and exit" to discard the calibration results and return to the Welcome screen.
Review Calibration History
1. Click "Review calibration history":
2. The CSV file viewer page appears (see below), which is time stamped, provides the operator ID, and
displays the "before" and "after" calibration image readings. The fields are not user editable in the viewer;
they are printable if desired.
LVS-7000 Operations Manual Version 5.6.X
LVS-7000 Operations Manual Version 5.6.X
Page 19 of 187