Wiring

6.1 Recommended cable entries (equivalents can be used)

6.1
Recommended cable entries (equivalents can be used)
General purpose cable entry: M20 (A5E03252531) and ½" NPT (A5E03252530)
1. Strip cable and expose braided shield.
2. Feed cable through dome nut and clamping insert. Fold braided shield over clamping insert.
Make sure that braided shield overlaps the O-ring by 3/32" (2 mm) and covers the entry 360
degrees.
3. Push clamping insert into body and tighten dome nut. Assemble into housing.