6. LOGFINISH
Resets the log to act as though the newest entry in the log has already been reported.
Syntax
LOGFINISH
Example
see LOGENTRY
7. LOGSTART
Resets the log iterator to the oldest entry in the log.
Syntax
LOGSTART
Example
see LOGENTRY
8. ** M
Move the specified motor to the specified position. If the motor hits a limit sensor
during its movement, then the motor is reset to its initial position and the requested
motor position is again moved to. This command requires configuration variables. See
the 10.31.7 for examples
Syntax
M,<m>,<s>
Parameters
The <m> identifies the motor to move. See the section on the I command for possible
values.
The <s> defines the motion required. If <s> is positive the motor moves to the given
position. If <s> is negative, the motor moves |<s>| steps backward and resets the 'zero'
position of the motor to be this new position.
Example
M, 4, 256
Moves the translucent ground quartz diffuser on filterwheel 1 into the optical path.
9. ** O
Transmit to the controller all photon count data accumulated by the most recent R
command, then zero the photon count accumulators. This command requires
configuration variables.
Syntax
O
Response
<c0>,<c1>,<c2>,...,<cn>
Each count number returned is in the form of a decimal value using ASCII characters.
Each number occupies exactly nine characters and the value is right-justified and
padded with spaces. The last value in the list does not have a comma or space but a
carriage return as expected.