HG Calibration Data
Example
Name
Hg
Type of measurement
12:10:22
Time of measurement
.9995
Correlation value
287.1829
Calculated micrometer step
287
Micrometer set to this step #
190255
Peak intensity of the HG scan
28
Temperature (deg. C)
Correlation Value:
Standard Lamp Test Data
#
Example
1
Sl
2
A
3
0
4
737.41
5
0
6
6
7
20
8
625382
9
43
10
644575
11
710539
12
839228
13
914419
14
981043
15
Rat
16
1543
17
1116
18
385
19
300
These lines are repeated for a total of 7 sets of data per SL test. This data is then averaged in the
following summary.
66
- The corelation between the stored and measured spectra.
- Calculated Micrometer Step number
- The micrometer position of the HG peak
Name
type of measurement
Filter
ND filter position of filterwheel #2 (in steps)
time – minutes since 00:00 hrs
Lower slit mask position (1=dark)
Upper slit mask position (6=slit 5)
# of cycles
raw counts wavelength #0
dark count
raw counts wavelength #1
raw counts wavelength #2
raw counts wavelength #3
raw counts wavelength #4
raw counts wavelength #5
ratios header
single ratio #1 MS(4)
single ratio #2 MS(5)
single ratio #3 MS(6)
single ratio #4 MS(7)