Enter 'y' if you want to update the calibration value; type 'n' if you feel the test was unsatisfactory
for any reason or if the value hasn't changed. A timestamped calibration record will be printed:
The Azimuth Tracker then returns to the solar azimuth and the previously displayed menu will
appear.

THE HUMIDITY SENSOR TEST

The Honeywell HIH-3605-A-CP sensor is a monolithic IC humidity sensor that provides a
proportional voltage output to relative humidity. The sensor is buffered by an Op Amp and the
output is connected to one of the analog input channels of the A/D converter. A temperature sensor
is located near the humidity sensor to provide temperature compensation for the device. The two
measured values, relative humidity and temperature allow the absolute humidity to be calculated.
The humidity sensor is supplied with a sensor specific data printout. The values of the data printout
are entered into a formula, which calculates the moisture content inside the instrument.
The Calibration confirmation test consists of placing the sensor in a sealed container with a variety
of water saturated salts that result in known, and accurate, RH values.
NaCl solution is 75.3%; the measured RH using the above formulas is 75.7%
MgCl
solution is 32.8%; the measured RH using the above formulas is 32.6%
2
LiCl solution is 11.3%; the measured RH using the above formulas is 12.3%
Dry N2 is 0%; the measured RH using the above formulas is 0.07%
These results show that the sensor is well within the range of its specifications
Method:
Attach the data printout from the humidity sensor container to this document.
Enter the values of origin and slope into the configuration file of the instrument.
Confirm normal operation of the instrument through the use of the routines TE and AP. Refer to the
operators manual for specific use of these routines.
Model: IH-3605B
Wafer: thunder
Pin 3:
+5 VOLTS PRECISION SUPPLY
Pin 2:
E+ SIGNAL OUTPUT
Pin 1:
NEGATIVE or POWER COMMON
HYCAL Sensing Products
Honeywell Inc.
9C Founders Blvd
El Paso TX 79906
Calculated Values at 5V:
Vout @0%=0.868 @75.3%=3.050
NaCl solution is 75.3%, the measured relative humidity is
MgCl
solution is 32.8%, the measured relative humidity is
2
LiCl solution is 11.3%, the measured relative humidity using is
Dry nitrogen solution is 0%, the measured relative humidity is
108
AZ STEPS/REV = 14675 AT HH:MM:SS
Channel:
189
MRP: thunder
Linear output for 2% RH accy @25C:
Zero offset = 0.868 V
Slope = 28.981 mV / %RH
RH = (Vout – 0.868) / 0.290
Radiometric response for 0 to 100%RH:
Vout = Vsupply * (0.1736 to 0.7533)
MKIII OPERATOR'S MANUAL
APPENDIX F FACTORY TESTS
File 98080318
%
%
%
%