OP_ST.nnn - Operating State File
The operating state file controls many of the operating parameters of the Brewer. The left column
lists the actual value written in the OP_ST.nnn file. The middle column is the BASIC variable name
used in the Brewer software to contain this value, and the right column is a description of the
value's meaning.
#
Sample
1
046
2
\BDATA
3
ICFjjjyy
4
ZSFjjjyy
5
DCFjjjyy
6
UVRjjjyy
7
01
8
01
9
99
10 Saskatoon LO$
11 52.108
12 106.713
13 960
14 1.8
15 215
16 0
17 14689
18 1
19 1
20 1
21 1
22 1
23 1
24 1
25 1
26 1
27 1
28 1
29 0
30 1
31 1
32 0
33 skc/menu
34 O
3
35 UMKNO2
82
SW
Explanation
Variable
NO$
Brewer ID number
DD$
Data Directory
ICF$
instrument constants file
ZSF$
zenith sky coefficients file
DCF$
dispersion constants file
UVR$
UV response file
DA$
Current day
MO$
Current month
YE$
Current year
Location of Brewer instrument
L1$
Latitude of instrument
L2$
Longitude of instrument
L3$
Average climatic station pressure (millibars)
TE$
Voltage representation of Brewer temperature
NC%
Azimuth north correction
HC%
Zenith horizon correction
SR%
Azimuth steps per revolution
Q1%
Zenith drive motor
Q2%
Azimuth drive motor
Q3%
Iris drive motor
Q4%
Filterwheel #1 drive motor
Q5%
Filterwheel #2 drive motor
Q6%
Clock board
Q7%
A/D board
Q8%
UVB port
Q9%
Filterwheel #3 drive motor
Q10%
New or old temperature circuit. Set to 1 for a new
temperature circuit and 0 for an old circuit.
Q11%
Second film polarizer
Q12%
Set to 1 to enable NOBREW operation
Q13%
Wide HG slit present. Always set this to 1.
Q14%
New Brewer electronics board
Q15%
Humidity Sensor
DI$
Schedule or menu indicator
MDD$
Mode
SK$
Schedule name
MKIII OPERATOR'S MANUAL