Kipp & Zonen Brewer Mk III Operator's Manual page 87

Table of Contents

Advertisement

UVRJJJYY.nnn
First column: wavelength in Angstrom.
Second column is responsivity in counts/mW/m
ZSFJJJYY.nnn (ZSFVAL)- Zenith Sky Constants
Zenith Sky constants are used in the ZS ozone calculations, and are Location/Brewer dependent.
They are derived by making a comparison of near simultaneous DS and ZS measurements over a
wide range of mu and ozone values (usually many months).
The values supplied in ZSFVAL.nnn are for a Brewer #035 operating in Toronto, Canada, and will
produce results which are accurate to within 5% or so. These values can be used until a new set
can be derived for the new site.
Example
Name
Zeni
Header
-.0064
Coefficient #1
-.01968
Coefficient #2
.01654
Coefficient #3
.194706
Coefficient #4
.280512
Coefficient #5
-.061317
Coefficient #6
-.490686
Coefficient #7
.456243
Coefficient #8
-.045191
Coefficient #9
84
2
/nm.
MKIII OPERATOR'S MANUAL

Advertisement

Table of Contents
loading