Chapter 6 Troubleshooting

Error Messages

An error message indicates an unusual condition. WARNING and FAULT error messages are a
result of exceeding one of the bath's Alarm limit settings (see Chapter 4), exceeding a sensor
factory preset safety value or a safety switch is activated. With either message the alarm, if
enabled, will sound.
In the case of a WARNING the bath, if running, will continue to run. Touch
message clears and, if enabled, silence the alarm. If the limit was only temporarily exceeded the
message will not reappear.
In the case of a FAULT the bath will shut down but the screen will continue to flash the message.
x
Touch
to clear the screen and, if enabled, silence the alarm. Once the cause of the shut
down is identified and corrected, restart the bath. If the cause of the fault was not corrected the
message will reappear.