UV Lamp Installation -
Models D8611 & D7401
Thermo Scientific

Installation

Section 6
See Figure 6-1 for details.
1. Disconnect the power cord from the unit.
2. Remove the screws securing the Easypure II top cover.
3. Remove the top cover by lifting straight up.
4. Locate the UV oxidation chamber and pull the top black plastic cover
off. Do not pull on the cable.
5. Remove the UV lamp from its packaging. DO NOT TOUCH THE
GLASS PORTION OF THE LAMP! It is recommended that lint-free
gloves be worn when handling the lamp. The glass portion must be
free of fingerprints, perspiration, etc. Even a light coating of
perspiration will reduce the effectiveness of the lamp. If the glass
portion of the lamp is touched, clean it with a damp, lint-free cloth.
Use isopropyl alcohol as required.
6. Connect the UV lamp to the receptacle in the black plastic cover. Slide
the lamp completely into the chamber and replace the black plastic
cover on the UV Chamber.
Figure 6-1. UV Lamp Installation
Black Plastic Cover
Receptacle
UV Lamp
UV Chamber
EASYpure® II
6-1