Replacement Procedures

Replacing the Transducer (continued)

Step
1.
Hold the Transducer Adapter mounting
with one hand and the Transducer Cover
assembly with the opposite hand. Gently
push the Transducer into the Transducer
Adapter mounting.
2.
Insert the four (4) screws (1500-0097-00)
into the Transducer Adapter mounting
and torque the screws to 10 in-lbs. (1.15
N-m)
NOTE: There are two (2) screws
on either side of the
Transducer Adapter
Mounting.
3.
Install the Scan Head assembly
Clear Cast Cover.
8-10
Description
Invenia ABUS 2.0 – System Setup and Basic Service Manual
Illustration
4700-0043-00 Rev. 4