NOTE:
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
c. Enter a unique Name for display in the patient entry
screen for the user to select at the time an exam is
started. The name should have real meaning to the
users who will select it as part of performing exams.
Figure 3-20. Destination Device
d. Enter the AE Title, IP Address and Port.
e. Touch the Return button
Export Images Settings: Unless customer is planning on reading
the Invenia ABUS 2.0 images on PACS do not modify the default
settings.
To delete a destination
a. Touch the Configuration button
General tab touch the Destination button.
b. Select the Destination name to be deleted.
c. Touch the Trash button to delete the entry.
Customizing the Invenia ABUS 2.0
to save the changes.
, and on the
3-59