Components and Functions (Theory)

Controls, displays, and ports

User controls
5-24
ABUS operator controls include:
1. Buttons on the Scan Head handles
These control Start / Stop scan, and compression adjustment.
2. Controls on the touchscreen monitor screen
These provide control of system settings and scan parameters.
Refer to the User Manual for full discussion of this topic.
3. (User-operable) Power on/off switch.
This illuminated pushbutton switch on the side of the Dual Box
chassis cover is for operator use in normal daily operation. It
lights green when the system is on. There is also a power switch
with thermal circuit breaker.
Scan Head controls
The two handles on the Scan Head provide operator control
pushbuttons.
The left handle buttons are:
start scan (green button)
stop/abort scan (red button)
The right handle buttons are:
decrease Scan Head pressure (up arrow)
increase Scan Head pressure (down arrow)
Touchscreen
The Touchscreen is touch-sensitive and supports the system
software detecting touches and gestures. It also provides an
onscreen keyboard for entering patient data. For details refer to
the Invenia 2.0 2.0 Basic User Manual (P/N 4700-0041-00).
Operator power switch
Although the ABUS 2.0 has a main power switch (with breaker)
on the ACDC input box, the ABUS provides a separate Operator
Power Switch pushbutton on the side of the chassis. This is
intended to be used for daily operations. The status of this
pushbutton is sensed by the Motherboard.