GE Invenia ABUS 2.0 Technical Publication page 269

Table of Contents

Advertisement

Biological
Hazard
Task
Clean Console
Clean Scan Head Assembly
Clean Touch Screen
Clean Filter(s)
NOTE:
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Cleaning takes about 15 minutes to perform.
Refer to the Invenia ABUS 2.0 user documentation whenever
necessary.
Possible risk of infection. Do not handle soiled or contaminated
transducer and other components that have been in patient
contact. Follow appropriate cleaning and surface-level
disinfecting procedures before handling the equipment.
Table 10-2:
Clean the console with disinfecting wipes.
Clean the Scan Head Assembly using the cleaning procedure on
page 10-6.
1. Turn off the Touch Screen Display.
2. Using the LCD-approved cleaning kit, clean the Touch Screen.
3. Turn the Touch Screen back on and verify calibration is still accurate.
Clean the Invenia ABUS 2.0 air filter using the procedure on page 10-6.
The germicidal wipes listed below are approved for use with the
Invenia ABUS 2.0.
List of Approved Disinfecting Wipes:
For the following disinfecting wipes, only chemical compatibility was
tested—the efficacy of these wipes as germicidal agents was not
evaluated:
1. Sani-Cloth HB Germicidal Wipes - PDI
2. Sani-Cloth AF3 Germicidal Wipes - PDI
3. Asepti-Wipes II Germicidal Wipes - Ecolab
4. Protex Ultra-Wipes - Parker Labs
5. Oxycide Daily Disinfectant Wipe Cleaner - Ecolab
6. Distel High-Level Disinfectant Wipes - Tristel
For the following disinfecting wipes, C. Diff was tested and passed:
1. Dispatch Hospital Cleaner Disinfectant Towels with Bleach – Clorox
2. Clinell Universal Wipes – GAMA Healthcare Ltd
Cleaning
Invenia ABUS 2.0
Cleaning
Cleaning