Diagnostics/Troubleshooting
System Messages
Symptom
Blank Screen
If the system powers on with an error on the screen.
Mechanical Noise / Thumping
Multi-Axis Lock Stops Working
System Overheating
Image doesn't transfer from the Invenia ABUS 2.0
system
Lines in the Image
Intermittent Operation, System Lock-Ups, System
Slow-Downs, Overheating, Poor Image Quality
7-30
Table 7-3:
Invenia ABUS 2.0 Messages
Invenia ABUS 2.0 – System Setup and Basic Service Manual
Action to Take
Troubleshoot the following:
• Is the system plugged in?
• Is the system turned on and the power button lit
(green)?
• Does the transducer initialize (move across the
face of the Scan Head Assembly)?
Reboot the system, up to three (3) times. If the error
still appears, contact GE Service.
Contact GE Service.
DO NOT spray any spray or oil, or apply any grease
to any part of the system.
Contact GE Service.
Clean the filter.
CAUTION: DO NOT operate the system without an
air filter.
Confirm that the Workstation Network Indicator light
is green. Shut down both the Invenia ABUS 2.0 and
Workstation; then 1st reboot the Workstation, and
2nd reboot the Invenia ABUS 2.0.
Remove the transducer, then reverse it 180 degrees
and plug it back in.
If you still see lines on the image, but on the
opposite side of the screen, then the transducer is
defective.
If the lines are still on the same side of the screen,
then the issue is on the adcquisition board or the TIP
board.
In either case, contact GE Service.
Clean the system filter and let the system cool down
for 30 minutes to 1 hour. Keep the system a distance
of 18 inches (~0.46 m) from curtains. CAUTION: DO
NOT operate the system without a filter.
4700-0043-00 Rev. 4