Diagnostics/Troubleshooting
Error Code Identifications
Ultrasound Box (GEUS) error code examples:
7-46
Example 1
Error code 35006 is related to high temperature readings in the
Ultrasound Box. Since this problem could be caused by a
blocked or partially blocked air-filter, the user is instructed to
check the system air filter first.
If the system continues to overheat, the customer places a
service call.
The service team can easily search the Service Manual by the
error code and follow the troubleshooting steps: check the cMST
temperature monitoring readings and run the system
diagnostics available in the Service Desktop to determine if the
problem is caused by the Ultrasound Box fans or possible
temperature reading errors on the cMST board sensors.
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4