Error
Category
Code
11001
Acquisition
12001
Acq-scanhead
13001
Renderer
14001
System status
14002
System status
21001
ScannerApp Init
21002
ScannerApp Init
21003
ScannerApp Init
21004
ScannerApp Init
21005
ScannerApp Init
21006
ScannerApp Init
21007
ScannerApp Init
22001
ScannerApp USB
22002
ScannerApp USB
31000
GEUS Unknown
31001
GEUS Failure
31002
GEUS Failure
31001
GEUS Failure
32001
GEUS Info
32002
GEUS Info
33001
GEUS Service
7-48
Description
Failed to initialize acquisition (missing frontend JSON file,
possible transducer disconnection)
Scanhead id does not match.
Image renderer initialization failed.
GPU Max temperature exceeded
Power Max exceeded
Initialization success
Weight motor init home failed
Scan motor init home failed
Ball motor init home failed
Scanhead controller USB COM port not detected
Weight controller USB COM port not detected
Scanhead and Weight controller USB COM port not detected
Scanhead controller USB COM port not detected
Weight controller USB COM port not detected
Unknown GEUS Error Code received
GEUS fatal error requiring system shutdown
GEUS fatal error requiring system restart
GEUS fatal error DSP not detected requiring system restart
GEUS information error
GEUS information error requiring system restart
GEUS service type error information category