Replacing The Transducer - GE Invenia ABUS 2.0 Technical Publication

Table of Contents

Advertisement

Replacement Procedures
1.
Power down the system by pressing the Power button on the front cover.
2.
Turn off the switch on the Power Entry Module/Power Cord Socket, then remove the Power Cord.
Maintain control of the power cord according to local lockout tagout requirements.
3.
Remove the Scan Head
assembly Clear Cast Cover.
4.
Remove the Stabilizing Membrane.
5.
Remove the four (4) screws securing the
Transducer Cover assembly to the
Transducer adapter mounting.
NOTE: There are two (2) screws
on either side of the
Transducer Adapter
Mounting.
8-8

Replacing the Transducer

To remove the transducer
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4