GE Invenia ABUS 2.0 Technical Publication page 230

Table of Contents

Advertisement

Diagnostics/Troubleshooting
Hardware
problems
Software problems
System Message
[Patient Name] is already selected for this scan
Abort fininshed. Ready for next scan.
Aborted
Aborting
Are you sure you want to logout?
Dataset is not available
Imaging Engine Initializing....
Please check the scantype
Please input patient information.
Please mark all nipple positions
Please select a new patient
Please unlock Scan Head Assembly first
Please wait until processing is finished
Press compression button
Press Compression button or Start button
Press Scan Button on Scan Head to Start Scanning
Press Start button to start scan
7-36
A hardware is an error/malfunction in either the transducer or
the main unit.
A hardware issue may be an error/malfunction on a card.
Check if a newer software version is available. A software
update may include noise fixes. If needed, update the software.
The following messages may appear during the scan. Please
refer to the cause/action information provided when trying to
resolve the issue.
Table 7-5:
Invenia ABUS 2.0 Messages
Invenia ABUS 2.0 – System Setup and Basic Service Manual
Cause/Action to Take
You have already selected a patient for this scan.
You can start the next scan; the previous scan was
aborted.
The scan has been aborted.
The current scan is being ended.
Prompts the user to confirm that they intended to
stop using the system.
The selected patient's dataset is no longer available.
It cannot be re-sent to the workstation.
The system is in the process of starting up.
The user must select a scan type before scanning
can begin.
Prompts the user to fill in all patient information fields
prior to starting the scan.
The nipple positions on the scan must be set before
proceeding with further action on the system.
You need to select a new patient in order to begin
the scan.
Login to begin scanning.
The system is still processing the current task;
please wait to start the next tast.
Apply compression before starting the scan.
Apply compression, then press Start to begin the
scan.
Press Start on the Scan Head Assembly to begin the
scan.
Press Start on the Scan Head Assembly to begin the
scan.