Cleaning the handpiece

Cleaning steps:
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Soak all parts in the cleaning solution for, as a minimum, the time and the concentration specified by the manufacturer of the cleaning agent (ENZOL, 2 %, 10
min).
• Brush all lumens and surfaces with a suitable lumen / surface brush until all visible soiling is removed.
• Rinse through all product lumina (e.g. irrigation and aspiration connection) with the cleaning solution (ENZOL, 2 %), at least 5 times in the flow direction (no
back rinsing), using a disposable syringe (min. volume 50 ml).
• Submerge all components completely in deionized water.
• Rinse through all product lumina (e.g. irrigation and aspiration connection) with deionized water at least 3 times in the flow direction (no back rinsing), using a
disposable syringe (min. volume 50 ml), using the Easy Clean (if available).
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the product.
• Dry all components at room temperature.
• Place the handpiece in a sterilization bag.