CURS100 100 Ohm Current Shunt
Terminal Input Module
1.

Introduction

NOTE
Terminal input modules connect directly to the data logger input terminals to
provide completion resistors for resistive bridge measurements, voltage
dividers, and precision current shunts. The CURS100 converts a current signal
(for example, 4 to 20 mA) to a voltage that is measured by the data logger. The
100-ohm resistor used for the current shunt allows currents up to 50 mA to be
read on a ±5000 mV range (CR6, CR800, CR850, CR1000, CR1000X,
CR3000, CR5000, CR9000X, CR9000). The CR300 allows currents up to
25 mA with a –100 to +2500 mV range.
The CR6 (serial numbers greater than or equal to 7502),
CR300-series, and CR1000X dataloggers are able to do current
measurements directly. While compatible with the CURS100, the
CURS100 is not required to convert the current signal to a voltage.
For more information, see CRBasic Editor Help for the
CurrentSE() instruction.
FIGURE 1-1. CURS100 terminal input module