Electrical Requirements - GE LOGIQ C9 Series Technical Publication

Hide thumbs Also See for LOGIQ C9 Series:
Table of Contents

Advertisement

Site Preparations

Electrical requirements

General requirements
NOTE:
NOTE:
Electrical requirements for the Ultrasound system
Table 2-3:
Tolerance
Voltage
±10%
220-240 VAC
2-6
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
In the table below, the electrical specifications for the Ultrasound
system includes monitor and on board peripherals.
Electrical Specifications for LOGIQ C9 Series system
Power Consumption
Max. 950VA
Frequency
50/ 60HZ
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5