How to Setup and Use a DICOM Image Storage Service

1
In Utility -> Connectivity -> Service, from the Destination Device drop-down menu, select
the device on which the service resides. (This assumes you have already setup the device in
the Device tab in 'How to Setup and Use a DICOM Image Storage Service' on page 3-37 .
2
From the Slect Service Type to add drop-down menu, select DICOM Image Storage and
press Add.
3
Change the name of the service to one of your choosing.
4
Enter the AE Title and port Number of the service. AE Title is case-sensitive.
4
Save your changes and then press Verify. A "Smiley Face" indicates successful
communication with the service.
Note: If you get a successful Ping (Smiley Face) at the device level but not at the service
level, it is possible that the AE Ttitle or Port Number of the service settings are incorrect.
Ensure that these are correct then re-verify. Be sure that the service type (Store, Print, etc) is
correct and supported by the device.
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
An Image Storage Service provides a place to store patient and
exam data from LOGIQ C9 Series and corresponding images.
The Image Storage Service, or the device that hosts it, is often
called a Patient Archiving and Communication System (PACS).
Table 3-18:
Setup an Image Storage Service
Connectivity setup
3-37