Air Filter Cleaning - GE LOGIQ C9 Series Technical Publication

Hide thumbs Also See for LOGIQ C9 Series:
Table of Contents

Advertisement

Replacement Procedures

Air Filter Cleaning

CAUTION
Tools
Preparation
Needed Manpower
8-10
Clean the system's air filter to ensure that a clogged filter does
not cause the system to overheat and reduce system
performance and reliability. It is recommended the filters be
cleaned every two weeks, but the requirements will vary with
environment.
Be sure to lock the wheels before cleaning the air filters to
aviod injury by any unexpected movement of the system.
DO NOT operate the unit without the air filters in place.
Allow the air filters to dry throughly before re-installing them on
the unit.
None
Shut down the system and disconnect the power cord.
1 person, 5 minutes + travel