Required facility needs (continued)
NOTE:

Desirable features

LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound unit is only a conduit.
Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit's proposed
location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.5m (1.5 ft.) from the
wall for cooling
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
Power outlets for other medical equipment.
Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrsound system.
Clean and protected space to store probes (in their cases or
on a rack)
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see: 'Electrical requirements'
on page 2-6.
Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily
accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe
sheaths
Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and
proprietary manuals
Facility needs
2-15