When There's Too Much Leakage Current; Ac/Dc Fails; Chassis Fails - GE LOGIQ C9 Series Technical Publication

Hide thumbs Also See for LOGIQ C9 Series:
Table of Contents

Advertisement

Care and Maintenance
When there's too much leakage

AC/DC Fails

Chassis Fails

NOTE:
10-30
Where applicable, check the AC/DC adapter and its cable.
Replace a new one if any portion is defective.
Check the ground on the power cord and plug for continuity.
Ensure the ground is not broken, frayed, or intermittent. Replace
any defective part.
Where applicable, tighten all grounds. Ensure star washers are
under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other
wiring abnormalities. Notify the user or owner to correct any
deviations. As a work around, check the other outlets to see if
they could be used instead.
No outlet tester can detect the condition where the white neutral
wire and the green grounding wire are reversed. If later tests
indicate high leakage currents, this should be suspected as a
possible cause and the outlet wiring should be visually
inspected.
current ...
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5